Dosimetric and Radiobiological Evaluation of Dose Calculation Algorithms On Prostate Stereotactic Body Radiotherapy Using Conventional Flattened and Flattening-Filter-Free Beam

no image available
S Kang

S Kang*,1, J Chung2, K Eom2, J Lee3, T Suh1,The catholic university of Korea, Seoul, 2Seoul National University Bundang Hospital, 3Konkuk University Medical Center

Presentations

SU-F-T-545 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: The purpose of this study is to evaluate the dosimetric and radiobiological impact of Acuros XB (AXB) and Anisotropic Analytic Algorithm (AAA) dose calculation algorithms on prostate stereotactic body radiation therapy plans with both conventional flattened (FF) and flattening-filter free (FFF) modes.

Methods: For thirteen patients with prostate cancer, SBRT planning was performed using 10-MV photon beam with FF and FFF modes. The total dose prescribed to the PTV was 42.7 Gy in 7 fractions. All plans were initially calculated using AAA algorithm in Eclipse treatment planning system (11.0.34), and then were re-calculated using AXB with the same MUs and MLC files. The four types of plans for different algorithms and beam energies were compared in terms of homogeneity and conformity. To evaluate the radiobiological impact, the tumor control probability (TCP) and normal tissue complication probability (NTCP) calculations were performed.

Results: For PTV, both calculation algorithms and beam modes lead to comparable homogeneity and conformity. However, the averaged TCP values in AXB plans were always lower than in AAA plans with an average difference of 5.3% and 6.1% for 10-MV FFF and FF beam, respectively. In addition, the averaged NTCP values for organs at risk (OARs) were comparable.

Conclusion: This study showed that prostate SBRT plan were comparable dosimetric results with different dose calculation algorithms as well as delivery beam modes. For biological results, even though NTCP values for both calculation algorithms and beam modes were similar, AXB plans produced slightly lower TCP compared to the AAA plans.