Hypo-Fractionation with Intensity Modulated Proton Therapy for Unilateral Metallic Prosthesis Prostate Cancer Patients

no image available
S Rana

S Rana1*, Y Zheng2 , Y Zhang3 , Pokharel4 , C Cheng5 , S Park1 , (1) McLaren Proton Therapy Center, Karmanos Cancer Institute at McLaren-Flint, Flint, MI, (2) Procure Proton Therapy Center, Oklahoma City, OK, (3) University of Cincinnati Medical Center, Liberty Township,, OH, (4) 21st Century Oncology, Estero, FL, (5) Vantage Oncology, West Hills, CA

Presentations

SU-F-T-196 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: The purpose of this study is to investigate the dosimetric feasibility of hypo-fractionated intensity modulated proton therapy (IMPT) for unilateral metallic prosthesis prostate cancer patients based on proton collaborative group (PCG)-GU002-10 (NCT01230866) protocol criteria.

Methods: A total of five unilateral metallic prosthesis prostate cancer cases were included in this retrospective study. For each case, IMPT plans were generated for treatment to be delivered with 7.6 Gy[RBE] per fraction in 5 fractions per week for a total dose of 38 Gy(RBE). Each plan was generated using two anterior-oblique beams and one lateral beam. Treatment plans were optimized with an objective meeting PCG-GU002-10 (NCT01230866) protocol criteria: (i) planning target volume (PTV): D99.5% > 36.1 Gy[RBE], (ii) rectum: V24 < 35%, V33.6 < 10%, (iii) bladder: V39 < 8 cc, and (iv) femoral head: V23 < 1cc.

Results: All five cases satisfied PTV D99.5% (average=36.82 Gy[RBE]; range, 36.36 – 37.13 Gy[RBE]). PTV D95% ranged from 36.66 Gy[RBE] to 38.65 Gy[RBE] and PTV V100 ranged from 95.47% to 97.95%. For the rectum, V24 was less than 35% (average=14.07 Gy[RBE]; range, 6.22 – 18.42%, whereas V33.6 Gy[RBE] was less than 10% (average=6.83; range, 3.06 – 9.15%). Rectal mean dose ranged from 4.22 Gy[RBE] to 9.97 Gy[RBE]. For the bladder, V39 was found to be less than 8 cc (average=3.69 cc; range, 0.19 - 7.68 cc). Bladder mean dose ranged from 4.22 Gy[RBE] to 18.83 Gy[RBE]. For the femoral head, V23 was 0 in all five cases.

Conclusion: All five unilateral metallic prosthesis prostate cancer IMPT plans generated with one lateral and two anterior-oblique beams satisfied the dosimetric criteria of PCG-GU002-10 (NCT01230866) protocol.