Study of Robustness Analysis Method of Multiple Field Optimized IMPT Plans for Head & Neck Patients


Y Li

Y Li1*, X Wang1 , H Li1 , J Wang1 , X Zhu1 , Y Li2 , S Frank1 , X Zhang1 , (1) UT MD Anderson Cancer Center, Houston, TX, (2) Varian Medical Systems, Houston, TX

Presentations

SU-F-T-192 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:Proton therapy is more sensitive to uncertainties than photon treatments due to protons’ finite range depending on the tissue density. Worst case scenario (WCS) method originally proposed by Lomax has been adopted in our institute for robustness analysis of IMPT plans. This work demonstrates that WCS method is sufficient enough to take into account of the uncertainties which could be encountered during daily clinical treatment.

Methods:A fast and approximate dose calculation method is developed to calculate the dose for the IMPT plan under different setup and range uncertainties. Effects of two factors, inversed square factor and range uncertainty, are explored. WCS robustness analysis method was evaluated using this fast dose calculation method. The worst-case dose distribution was generated by shifting isocenter by 3 mm along x,y and z directions and modifying stopping power ratios by ±3.5%. 1000 randomly perturbed cases in proton range and x, y ,z directions were created and the corresponding dose distributions were calculated using this approximated method. DVH and dosimetric indexes of all 1000 perturbed cases were calculated and compared with the result using worst case scenario method.

Results:The distributions of dosimetric indexes of 1000 perturbed cases were generated and compared with the results using worst case scenario. For D95 of CTVs, at least 97% of 1000 perturbed cases show higher values than the one of worst case scenario. For D5 of CTVs, at least 98% of perturbed cases have lower values than worst case scenario.

Conclusion:By extensively calculating the dose distributions under random uncertainties, WCS method was verified to be reliable in evaluating the robustness level of MFO IMPT plans of H&N patients. The extensively sampling approach using fast approximated method could be used in evaluating the effects of different factors on the robustness level of IMPT plans in the future.