Multi-Lesion Cranial SRS VMAT Plan Quality

no image available
A Ballangrud

A Ballangrud*, L Kuo , L Happersett , S Lim , X Li , K Beal , Y Yamada , T LoSasso , J Mechalakos , Memorial Sloan-Kettering Cancer Center, New York, NY

Presentations

SU-F-T-613 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: Cranial SRS VMAT plans must have steep dose gradient around each target to reduce dose to normal brain. This study reports on the correlation between gradient index (GI=V50%/V100%), target size and target dose heterogeneity index (HI=PTV Dmax/prescription dose) for multi-lesion cranial SRS VMAT plans.

Methods: VMAT plans for 10 cranial cases with 3 to 6 lesions (total 39 lesions) generated in Varian Eclipse V11.0.47 with a fine-tuned AAA beam model and 0.125 cm dose grid were analyzed. One or two iso centers were used depending on the spatial distribution of lesions. Two to nine coplanar and non-coplanar arcs were used per isocenter. Conformity index (CI= V100%/VPTV), HI, and GI were determined for each lesion. Dose to critical structures were recorded.

Results: Lesion size ranged from 0.05–11.00 cm3. HI ranged from 1.2-1.4, CI ranged from 1.0–2.8 and GI from 3.1–8.4. Maximum dose to brainstem, chiasm, lenses, optic nerves and eyes ranged from 120-1946 cGy, 47-463 cGy, 9-121 cGy, 14-512 cGy, and 17-294 cGy, respectively. Brain minus PTV (Brain-PTV) V7Gy was in the range 1.1–6.5%, and Brain-PTV Dmean was in the range 94–324 cGy.

Conclusion: This work shows that a GI < 5 can be achieved for lesions > 0.4cc. For smaller lesions, GI increases rapidly. GI is lower when HI is increased. Based on this study, recommend HI is 1.4, and recommended GI is for volumes <0.1cc GI<9, 0.1-0.4cc GI<6, 0.4-0.1.0cc GI<5, and for volumes >1.0cc GI<4. CI is < 1.3 for all lesions except for targets < 0.1cc. Cranial SRS VMAT plans must be optimized to lower the GI to reduce the dose to normal brain tissue.