4Dand 5D Proton Dose Evaluation with Monte Carlo

no image available
U Titt

U Titt*, D Mirkovic , P Yepes , A Liu , C Peeler , S Randenyia , R Mohan , UT MD Anderson Cancer Center, Houston, TX

Presentations

SU-F-T-122 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: We evaluated uncertainties in therapeutic proton doses of a lung treatment, taking into account intra-fractional geometry changes, such as breathing, and inter-fractional changes, such as tumor shrinkage and weight loss.

Methods:A Monte Carlo study was performed using four dimensional CT image sets (4DCTs) and weekly repeat imaging (5DCTs) to compute fixed RBE (1.1) and variable RBE weighted dose in an actual lung treatment geometry. The MC2 Monte Carlo system was employed to simulate proton energy deposition and LET distributions according to a thoracic cancer treatment plan developed with a 3D-CT in a commercial treatment planning system, as well as in each of the phases of 4DCT sets which were recorded weekly throughout the course of the treatment. A cumulative dose distribution in relevant structures was computed and compared to the predictions of the treatment planning system.

Results:Using the Monte Carlo method, dose deposition estimates with the lowest possible uncertainties were produced. Comparison with treatment planning predictions indicates that significant uncertainties may be associated with therapeutic lung dose prediction from treatment planning systems, depending on the magnitude of inter- and intra-fractional geometry changes.

Conclusion:As this is just a case study, a more systematic investigation accounting for a cohort of patients is warranted; however, this is less practical because Monte Carlo simulations of such cases require enormous computational resources. Hence our study and any future case studies may serve as validation/benchmarking data for faster dose prediction engines, such as the track repeating algorithm, FDC.


Contact Email: