Encrypted login | home

Program Information

On the Determination of Equivalent Squares for Rectangular Small MV Photon Fields


O Sauer

OA Sauer*, S Wegener , F Exner , University of Wuerzburg, Wuerzburg, Germany

Presentations

SU-F-T-408 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: It is common practice to tabulate dosimetric data like output factors, scatter factors and detector signal correction factors for a set of square fields. In order to get the data for an arbitrary field, it is mapped to an equivalent square, having the same scatter as the field of interest. For rectangular fields both, tabulated data and empiric formula exist. We tested the applicability of such rules for very small fields.

Methods: Using the Monte-Carlo method (EGSnrc-doseRZ), the dose to a point in 10cm depth in water was calculated for cylindrical impinging fluence distributions. Radii were from 0.5mm to 11.5mm with 1mm thickness of the rings. Different photon energies were investigated. With these data a matrix was constructed assigning the amount of dose to the field center to each matrix element. By summing up the elements belonging to a certain field, the dose for an arbitrary point in 10cm depth could be determined. This was done for rectangles up to 21mm side length. Comparing the dose to square field results, equivalent squares could be assigned. The results were compared to using the geometrical mean and the 4Xperimeter/area rule.

Results: For side length differences less than 2mm, the difference between all methods was in general less than 0.2mm. For more elongated fields, relevant differences of more than 1mm and up to 3mm for the fields investigated occurred. The mean square side length calculated from both empiric formulas fitted much better, deviating hardly more than 1mm and for the very elongated fields only.

Conclusion: For small rectangular photon fields, deviating only moderately from square both investigated empiric methods are sufficiently accurate. As the deviations often differ regarding their sign, using the mean improves the accuracy and the useable elongation range. For ratios larger than 2, Monte-Carlo generated data are recommended.


Funding Support, Disclosures, and Conflict of Interest: SW is funded by Deutsche Forschungsgemeinschaft (SA481/10-1)


Contact Email: