Dosimetric Comparison of Coplanar and Non-Coplanar IMRT Plans for Peripheral Lung Lesion

no image available
J Kang

J Kang1*, S Zhang1 , S Philbrook1 , S Paul1 , B Wang2 , (1) Abington Memorial Hospital, Abington, PA, (2) Fox Chase Cancer Center, Philadelphia, PA

Presentations

SU-F-T-416 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:The purpose of this study was to compare dosimetric parameters of treatment plans between coplanar and non-coplanar techniques for treating peripheral lung lesions.

Methods:The planning CT scans of 6 patients in supine positions were used in this study. The size of the PTV ranges from 163 c.c. to 782 c.c.. The locations of PTV are mostly at the peripheral of Lung, some spreading to the mediastinum. For each patient, we generated two IMRT plans, one with and the other without non-coplanar beams. The non-coplanar beams were carefully selected so that the beams would never exit patient bodies through the contralateral lung. The IMRT plans were generated with Pinnacle 9.8 treatment planning software. The IMRT optimization objectives were kept the same for the corresponding pairs of plans. All plans were normalized such that 95% of PTV receives the prescription dose (full dose).

Results:The conformity index (mean±standard deviation of the mean) is 1.49±0.14 and 1.58±0.23 for the coplanar and noncoplanar plans, respectively. The heterogeneity index (mean±standard deviation of the mean) is 7.74 ±2.33 and 6.34±1.40 for the coplanar and non-coplanar plans, respectively. The maximum heart dose is 60.94±6.22 and 60.42±7.21 Gy, and mean heart dose is 10.22 ±7.57, 9.07 ±6.32 Gy, for the coplanar and non-coplanar plans, respectively. The ipsilateral lung V20 is 48.0%±2.4% and 47.5%±3.3%, and V5 is 68.2%±10.0% and 69.1%±7.3%, for the coplanar and noncoplanar plans, respectively. Furthermore, with the non-coplanar beam arrangement, the contralateral lung V20 was reduced from 3.3%±3.7% to 1.3%±0.8%, and the contralateral Lung V5 is reduced significantly from 65.6%±9.3% to 33.5%±20.9% (p value =0.008).


Conclusion:The IMRT plans with non-coplanar beam arrangement could reduce the exit dose to the contralateral lung, and therefore reduce the contralateral lung V5 significantly. This method is especially helpful while the lung lesion doesn’t have a symmetric shape.