B Ganesan

B ganesan1*, A Prakasarao2 , T Palraj3 , R RAI4 , G Singaravelu5 , (1) ,,,(2) ANNA UNIVERSITY, Chennai, TamilNadu, (3) Dr.Rai Memorial Cancer Institute, Chennai, TamilNadu, (4) Dr.Rai Memorial Cancer Institute, Chennai, TamilNadu, (5) ANNA UNIVERSITY, Chennai, TamilNadu

Presentations

SU-F-T-474 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
The use of mega voltage gamma and x-ray sources with their skin sparring qualities in radiation therapy has been a boon in relieving patient discomfort and allowing high tumor doses to be given with fewer restrictions due to radiation effects in the skin. However, high doses given to deep tumors may require careful consideration of dose distribution in the buildup region in order to avoid irreparable damage to the skin.

Methods:
To measure the perturbation of MOSFET detector in Co60,6MV and 15MV the detector was placed on the surface of the phantom covered with the brass build up cap. To measure the effect of temperature the MOSFET detector was kept on the surface of hot water polythene container and the radiation was delivere .In order to measure the sensitivity variation with accumulated dose Measurements were taken by delivering the dose of 200 cGy to MOSFET until the MOSFET absorbed dose comes to 20,000 cGy

Results:
the Measurement was performed by positioning the bare MOSFET and MOSFET with brass build up cap on the top surface of the solid water phantom for various field sizes in order to find whether there is any attenuation caused in the dose distribution.The response of MOSFET was monitored for temperature ranging from 42 degree C to 22 degree C.The integrated dose dependence of MOSFET dosimeter sensitivity over different energy is not well characterized.This work investigates the dual-bias MOSFET dosimeter sensitivity response to 6 MV and 15 MV beams.
Conclusion:
From this study it is observed that unlike diode, bare MOSFET does not perturb the radiation field.. It is observed that the build-up influences the temperature dependency of MOSFET and causes some uncertainty in the readings.In the case of sensitivity variation with accumulated dose MOSFET showed higher sensitivity with dose accumulation for both the energies.