Patient-Specific Point Dose Measurement Using Fiber Optic Radiation Sensor Using Cerenkov Radiation for Proton Therapeutic Beam

no image available
J Son

J Son1, 4*, M Kim2 , M Yoon3 , D Shin4 , (1) Korea University, Seoul, Seoul, (2) Dongnam Institute of Radiological & Medical Sciences, Busan, Busan, (3) Korea University, Seoul, Korea, (4) National Cancer Center, Goyang-si,,

Presentations

SU-F-T-174 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: A fiber-optic radiation sensor using Cerenkov radiation (FOCR) has been widely studied for use as a dosimeter for proton therapeutic beam. We developed the FOCR, and it applied to patient-specific point dose measurement in order to evaluate the effectiveness of the FOCR system for proton therapy QA.

Methods: Calibration of FOCR was performed with an ionization chamber whose absolute doses were determined according to the IAEA TRS-398 protocol. To determine the calibration curve, the FOCR was irradiated perpendicularly to the proton beam at the 13 dose levels steps. We selected five actual patient treatment plans performed at proton therapy center and compared the resulting FOCR measurements with the ionization chamber measurements.

Results: The Cerenkov light yield of the FOCR increases linearly with as the dose measured using the ionization chamber increases from 0 cGy to 500 cGy. The results indicate that the fitting curve is linear, suggesting that dose measurement based on the light yield of the FOCR is possible. The results of proton radiation dose QA performed using the FOCR for 10 proton fields and five patients are good agreement with an ionization chamber.

Conclusion: We carried out the patient QA using the FOCR for proton therapeutic beam and evaluated the effectiveness of the FOCR as a proton therapy QA tool. Our results indicate that the FOCR is suitable for use in patient QA of clinical proton beams.