Infant Total Skin Electron Therapy Using Five Fields Technique


H Saleh

H Saleh1*, T Howlin2 , V Massey3 , University of Kansas Hospital, Overland Park, KS

Presentations

SU-F-T-83 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: We were presented with a 9 month old boy with Relapsed Acute Myeloid Leukemia (AML) involving the skin. The plan was to treat the entire skin using 6 MeV electrons with the infant under complete anesthesia. The purpose of this work is to commission the 6 MeV electron beam and to develop a technique that can be used to deliver total skin dose to infants with minimal patient immobilization.

Methods: A baby mannequin phantom that mimics the child’s length was used to determine the best technique to treat the infant. The 76 cm long phantom was placed on the floor. The phantom was placed in four unique immobilization devices to simulate four different treatment positions (anterior, posterior, left lateral and right lateral). Radiochromic films were used to determine beam profile in both axial and radial directions, and percent depth dose (PDD). Absolute calibration of the machine output at 214 cm distance was measured using an Exradin A11 parallel-plate ion chamber. A 1.0 cm plexiglass scatter plate was inserted in the collimator. Mosfet dosimeters were used for dose verification for phantom and and patient.

Results: At 214 cm source to surface distance (SSD) using gantry angle of + 20o from vertical beam flatness is + 10% in the radial direction over a region of 70 cm and + 4% in the axial direction over 60 cm. A five field arrangement was determined to optimally deliver the desired dose with > 90% uniformity. The fifth field was used to boost the head vertex.

Conclusion:It is possible to treat sedated infants with total skin dose using five positions. Four positions were enough to cover the body and the fifth position boosts the vertex of the head. All fractions can be reproduced accurately daily because of the patient’s stable immobilization.


Contact Email: