Encrypted login | home

Program Information

ECG-Gated CT Reconstruction for a C-Arm Inverse Geometry X-Ray System


J Slagowski

JM Slagowski*, DAP Dunkerley , MA Speidel , University of Wisconsin - Madison, Madison, WI

Presentations

MO-DE-207A-6 (Monday, August 1, 2016) 1:45 PM - 3:45 PM Room: 207A


Purpose: To obtain ECG-gated CT images from truncated projection data acquired with a C-arm based inverse geometry fluoroscopy system, for the purpose of cardiac chamber mapping in interventional procedures.

Methods: Scanning-beam digital x-ray (SBDX) is an inverse geometry fluoroscopy system with a scanned multisource x-ray tube and a photon-counting detector mounted to a C-arm. In the proposed method, SBDX short-scan rotational acquisition is performed followed by inverse geometry CT (IGCT) reconstruction and segmentation of contrast-enhanced objects. The prior image constrained compressed sensing (PICCS) framework was adapted for IGCT reconstruction to mitigate artifacts arising from data truncation and angular undersampling due to cardiac gating. The performance of the reconstruction algorithm was evaluated in numerical simulations of truncated and non-truncated thorax phantoms containing a dynamic ellipsoid to represent a moving cardiac chamber. The eccentricity of the ellipsoid was varied at frequencies from 1-1.5 Hz. Projection data were retrospectively sorted into 13 cardiac phases. Each phase was reconstructed using IGCT-PICCS, with a nongated gridded FBP (gFBP) prior image. Surface accuracy was determined using Dice similarity coefficient and a histogram of the point distances between the segmented surface and ground truth surface.

Results: The gated IGCT-PICCS algorithm improved surface accuracy and reduced streaking and truncation artifacts when compared to nongated gFBP. For the non-truncated thorax with 1.25 Hz motion, 99% of segmented surface points were within 0.3 mm of the 15 mm diameter ground truth ellipse, versus 1.0 mm for gFBP. For the truncated thorax phantom with a 40 mm diameter ellipse, IGCT-PICCS surface accuracy measured 0.3 mm versus 7.8 mm for gFBP. Dice similarity coefficient was 0.99-1.00 (IGCT-PICCS) versus 0.63-0.75 (gFBP) for intensity-based segmentation thresholds ranging from 25-75% maximum contrast.

Conclusions: The PICCS algorithm was successfully applied to reconstruct truncated IGCT projection data with angular undersampling resulting from simulated cardiac gating.


Funding Support, Disclosures, and Conflict of Interest: Research supported by the National Heart, Lung, and Blood Institute of the NIH under award number R01HL084022. The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIH.


Contact Email: