Encrypted login | home

Program Information

Vulnerabilities in the Gamma Metric


B Neal

B Neal*, J Siebers , University of Virginia Health System, Charlottesville, VA

Presentations

SU-G-BRB-16 (Sunday, July 31, 2016) 4:00 PM - 6:00 PM Room: Ballroom B


Purpose: To explore vulnerabilities in the gamma index metric that undermine its wide use as a radiation therapy quality assurance tool.

Methods: 2D test field pairs (images) are created specifically to achieve high gamma passing rates, but to also include gross errors by exploiting the distance-to-agreement and percent-passing components of the metric. The first set has no requirement of clinical practicality, but is intended to expose vulnerabilities. The second set exposes clinically realistic vulnerabilities. To circumvent limitations inherent to user-specific tuning of prediction algorithms to match measurements, digital test cases are manually constructed, thereby mimicking high-quality image prediction.

Results: With a 3 mm distance-to-agreement metric, changing field size by ±6 mm results in a gamma passing rate over 99%. For a uniform field, a lattice of passing points spaced 5 mm apart results in a passing rate of 100%. Exploiting the percent-passing component, a 10x10 cm² field can have a 95% passing rate when an 8 cm²=2.8x2.8 cm² highly out-of-tolerance (e.g. zero dose) square is missing from the comparison image. For clinically realistic vulnerabilities, an arc plan for which a 2D image is created can have a >95% passing rate solely due to agreement in the lateral spillage, with the failing 5% in the critical target region. A field with an integrated boost (e.g whole brain plus small metastases) could neglect the metastases entirely, yet still pass with a 95% threshold. All the failure modes described would be visually apparent on a gamma-map image.

Conclusion: The %gamma<1 metric has significant vulnerabilities. High passing rates can obscure critical faults in hypothetical and delivered radiation doses. Great caution should be used with gamma as a QA metric; users should inspect the gamma-map. Visual analysis of gamma-maps may be impractical for cine acquisition.


Contact Email: