An Automated Technique for Estimating Patient-Specific Regional Imparted Energy and Dose From TCM CT Exams Across 13 Protocols

no image available
J Sanders

J Sanders1*, X Tian2, P Segars3, J Boone4, E Samei5, (1) Duke University, Durham, NC, (2) Duke University, Durham, NC, (3) Duke University, Durham, NC, (4) UC Davis Medical Center, Sacramento, CA, (5) Duke University Medical Center, Durham, NC

Presentations

TU-H-207A-9 (Tuesday, August 2, 2016) 4:30 PM - 6:00 PM Room: 207A


Purpose: To develop an automated technique for estimating patient-specific regional imparted energy and dose from tube current modulated (TCM) computed tomography (CT) exams across a diverse set of head and body protocols.

Methods: A library of 58 adult computational anthropomorphic extended cardiac-torso (XCAT) phantoms were used to model a patient population. A validated Monte Carlo program was used to simulate TCM CT exams on the entire library of phantoms for three head and 10 body protocols. The net imparted energy to the phantoms, normalized by dose length product (DLP), and the net tissue mass in each of the scan regions were computed. A knowledgebase containing relationships between normalized imparted energy and scanned mass was established. An automated computer algorithm was written to estimate the scanned mass from actual clinical CT exams. The scanned mass estimate, DLP of the exam, and knowledgebase were used to estimate the imparted energy to the patient. The algorithm was tested on 20 chest and 20 abdominopelvic TCM CT exams.

Results: The normalized imparted energy increased with increasing kV for all protocols. However, the normalized imparted energy was relatively unaffected by the strength of the TCM. The average imparted energy was 681 ± 376 mJ for abdominopelvic exams and 274 ± 141 mJ for chest exams. Overall, the method was successful in providing patient-specific estimates of imparted energy for 98% of the cases tested.

Conclusion: Imparted energy normalized by DLP increased with increasing tube potential. However, the strength of the TCM did not have a significant effect on the net amount of energy deposited to tissue. The automated program can be implemented into the clinical workflow to provide estimates of regional imparted energy and dose across a diverse set of clinical protocols.