Validation of a Monte Carlo Radiographic Simulation


A Gearhart

A Gearhart*, D Carver , M Stabin , Vanderbilt University, Nashville, TN

Presentations

SU-F-I-79 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To validate a radiographic simulation in order to estimate patient dose due to clinically-used radiography protocols.

Methods: A Monte Carlo simulation was created to simulate a radiographic x-ray beam using GEANT4. Initial validation was performed according to a portion of TG 195. Computational NURBS-based phantoms were used simulate patients of varying ages and sizes. The deposited energy in the phantom is output by the simulation. The exposure in air from a clinically used radiography unit was measured at 100 cm for various tube potentials. 10 million photons were simulated with 1 cubic centimeter of air located 100 cm from the source, and the total absorbed dose was noted. The normalization factor was determined by taking a ratio of the measured dose in air to the simulated dose in air. Dose to individual voxels is calculated using the energy deposition map along with the voxelized and segmented phantom and the normalization factor. Finally, the effective dose is calculated using the ICRP methodology and tissue weighting factors.

Results: This radiography simulation allows for the calculation and visualization of the energy deposition map within a voxelized phantom. The ratio of exposure, measured using an ionization chamber, to air in the simulation was determined. Since the simulation output is calibrated to match the exposure of a given clinical radiographic x-ray tube, the dose map may be visualized. This will also allow for absorbed dose estimation in specific organs or tissues as well as a whole body effective dose estimation.

Conclusion: This work indicates that our Monte Carlo simulation may be used to estimate the radiation dose from clinical radiographic protocols. This will allow for an estimate of radiographic dose from various examinations without the use of traditional methods such as thermoluminescent dosimeters and body phantoms.


Contact Email: