Encrypted login | home

Program Information

CTDI Per KV at Phantom Center and Periphery: Comparison Between Major CT Manufacturers


R Al-Senan

R Al-Senan1*, O DEMIRKAYA2 , (1) Columbia University Medical Center, New York, NY, (2) King Faisal Specialist Hospital & Research Centre, Riyadh, Saudi Arabia

Presentations

SU-G-206-3 (Sunday, July 31, 2016) 4:00 PM - 6:00 PM Room: 206


Purpose: The purpose of this study was to: 1) compare scanners output by measuring normalized CTDIw (mGy/100mAs) in different CT makes and models and at different kV’s, and 2) quantify the relationship between kV and CTDI and compare this relationship between the different manufacturers.

Methods: Study included forty scanners of major CT manufacturers and of various models. Exposure was measured at center and 12 o’clock holes of head and body CTDI phantoms, at all available kV’s, and with the largest or second largest available collimation in each scanner. Average measured CTDI’s from each CT manufacturer were also plotted against kV and the fitting equation: CTDIw (normalized) = a.kVb was calculated. The power (b) value may be considered as an indicator of spectral filtration, which affects the degree of beam hardening. Also, HVLs were measured at several scanners.

Results: Results showed GE scanners, on average, had higher normalized CTDIw than those of Siemens and Philips, in both phantom sizes and at all kV’s. ANOVA statistic indicated the difference was statistically significant (p < 0.05). Comparison between Philips and Siemens, however, was not statistically significant. Curve fitting showed b values ranged from 2.4 to 2.9 (for Head periphery and center, respectively); and was about 2.8 for Body phantom periphery, and 3.2 at the center of Body phantom. Fitting equations (kV vs. CTDI) will be presented and discussed. GE’s CTDIw vs. HVL showed very strong correlation (r > 0.99).

Conclusion: Partial characterization of scanners output was performed which may be helpful in dose estimation to internal organs. The relatively higher output from GE scanners may be attributed to lower filtration. Work is still in progress to obtain CTDI values from other scanners as well as to measure their HVLs.


Contact Email: