Efficient Data Collection of Continuous 2D and Discrete Relative Dosimetric Data for Annual LINAC QA Using TrueBeam Developer Mode and a 1D Scanning Tank

no image available
N Knutson

N Knutson1,2,4*, M Schmidt1,2,4 , N Nguyen1,4 , M Belley1,2 , M Price1,2,3 , (1) Rhode Island Hospital, Providence , RI, (2) University of Rhode Island, Kingston, RI (3) Alpert Medical School of Brown University, Providence, RI, (4) University of Massachusetts Lowell, Lowell, MA,

Presentations

MO-FG-202-3 (Monday, August 1, 2016) 4:30 PM - 6:00 PM Room: 202


Purpose: To develop a method to exploit real-time dynamic machine and couch parameter control during linear accelerator (LINAC) beam delivery to facilitate efficient performance of TG-142 suggested, Annual LINAC QA tests.

Methods: Varian’s TrueBeam Developer Mode (Varian Medical Systems, Palo Alto, CA) facilitates control of Varian’s TrueBeam LINAC via instructions provided in Extensible Markup Language (XML) files. This allows machine and couch parameters to be varied dynamically, in real-time, during beam delivery. Custom XML files were created to allow for the collection of (1) continuous Tissue Maximum Ratios (TMRs), (2) beam profiles, and (3) continuous output factors using a 1D-scanning tank. TMRs were acquired by orienting an ionization chamber (IC) at isocenter (depth=25cm) and synchronizing a depth scan towards the water surface while lowering the couch at 1mm/s. For beam profiles, the couch was driven laterally and longitudinally while logging IC electrometer readings. Output factors (OFs) where collected by continually varying field sizes (4x4 to 30x30-cm²) at a constant speed of 6.66 mm/s. To validate measurements, comparisons were made to data collected using traditional methods (e.g. 1D or 3D tank).

Results: All data collecting using the proposed methods agreed with traditionally collected data (TMRs within 1%, OFs within 0.5% and beam profile agreement within 1% / 1mm) while taking less time to collect (factor of approximately 1/10) and with a finer sample resolution.

Conclusion: TrueBeam developer mode facilitates collection of continuous data with the same accuracy as traditionally collected data with a finer resolution in less time. Results demonstrate an order of magnitude increase in sampled resolution and an order of magnitude reduction in collection time compared to traditional acquisition methods (e.g. 3D scanning tank). We are currently extending this approach to perform other TG-142 tasks.