Development of Conversion Factors to Account for Different CTDI Phantom Setups On a Wide-Beam CT Scanner
D Zamora , J Moirano , K Kanal*, University Washington, Seattle, WA
Presentations
TU-H-207A-1 (Tuesday, August 2, 2016) 4:30 PM - 6:00 PM Room: 207A
Purpose: A fundamental measure performed during an annual physics CT evaluation confirms that system displayed CTDIvol nearly matches the independently measured value in phantom. For wide-beam (z-direction) CT scanners, AAPM Report 111 defined an ideal measurement method; however, the method often lacks practicality. The purpose of this preliminary study is to develop a set of conversion factors for a wide-beam CT scanner, relating the CTDIvol measured with a conventional setup (single CTDI phantom) versus the AAPM Report 111 approach (three abutting CTDI phantoms).
Methods: For both the body CTDI and head CTDI, two acquisition setups were used: A) conventional single phantom and B) triple phantom. Of primary concern were the larger nominal beam widths for which a standard CTDI phantom setup would not provide adequate scatter conditions. Nominal beam width (160 or 120 mm) and kVp (100, 120, 140) were modulated based on the underlying clinical protocol. Exposure measurements were taken using a CT pencil ion chamber in the center and 12 o’clock position, and CTDIvol was calculated with ‘nT’ limited to 100 mm. A conversion factor (CF) was calculated as the ratio of CTDIvol measured in setup B versus setup A.
Results: For body CTDI, the CF ranged from 1.04 up to 1.10, indicating a 4-10% difference between usage of one and three phantoms. For a nominal beam width of 160 mm, the CF did vary with selected kVp. For head CTDI at nominal beam widths of 120 and 160 mm, the CF was 1.00 and 1.05, respectively, independent of the kVp used (100, 120, and 140).
Conclusions: A clear understanding of the manufacturer method of estimating the displayed CTDIvol is important when interpreting annual test results, as the acquisition setup may lead to an error of up to 10%. With appropriately defined CF, single phantom use is feasible.