Is the Reproducibility with Respect to Bone of Tumor Position at Simulation for Breath Hold CT Scans Correlated to the Reproducibility for Multiple Breath Hold CBCTs at Treatment in SBRT Thoracic Patients?

no image available
J Pollard

J Pollard*, S Prajapati , S Gao , P Nitsch , R Sadagopan , X Wang , P Balter , UT MD Anderson Cancer Center, Houston, TX

Presentations

SU-G-JeP3-15 (Sunday, July 31, 2016) 5:00 PM - 5:30 PM Room: ePoster Theater


Purpose:
To evaluate correlation between the reproducibility of tumor position under feedback guided voluntary deep inspiration breath hold gating at simulation and at treatment.

Methods:
All patients treated with breath hold (BH) have 3-6 BH CTs taken at simulation (sim). In addition, if the relationship between the tumor and nearby bony anatomy on treatment BH CT(or CBCT) is found to be greater than 5 mm different at treatment than it was at sim, a repeat BH CT is taken before treatment. We retrospectively analyzed the sim CTs for 19 patients who received BH SBRT lung treatments and had repeat BH CT on treatment. We evaluated the reproducibility of the tumor position during the simulation CTs and compared this to the reproducibility of the tumor position on the repeat treatment CT with our in-house CT alignment software (CT-Assisted Targeting for Radiotherapy).

Results:
Comparing the tumor position for multiple simulation BH CTs, we calculated: maximum difference (max) = 0.69cm; average difference (x) = 0.28cm; standard deviation (σ) = 0.18cm. Comparing the repeat BH CBCTs on treatment days we calculated: max = 0.44cm; x = 0.16cm; σ = 0.22cm. We also found that for 95% of our BH cases, the absolute variation in tumor position within the same imaging day was within 5mm of the range at the time of simulation and treatment. We found that 75% of the BH cases had less residual tumor motion on treatment days than at simulation.

Conclusion:
This suggests that a GTV contour based upon the residual tumor motion in multiple BH datasets plus 2 mm margin should be sufficient to cover the full range of residual tumor motion on treatment days.