A Backscatter Point Spread Function for Entrance Skin Dose Determination

no image available
S Vijayan

S Vijayan*, Z Xiong , A Shankar , S Rudin , D Bednarek ,Toshiba Stroke and Vascular Research Center, University at Buffalo (SUNY), Buffalo, NY

Presentations

TU-D-209-2 (Tuesday, August 2, 2016) 11:00 AM - 12:15 PM Room: 209


Purpose:To determine the distribution of backscattered radiation to the skin resulting from a non-uniform distribution of primary radiation through convolution with a backscatter point spread function (PSF).

Methods:A backscatter PSF is determined using Monte Carlo simulation of a 1 mm primary beam incident on a 30 x 30 cm x 20 cm thick PMMA phantom using EGSnrc software. A primary profile is similarly obtained without the phantom and the difference from the total provides the backscatter profile. This scatter PSF characterizes the backscatter spread for a “point” primary interaction and can be convolved with the entrance primary dose distribution to obtain the total entrance skin dose. The backscatter PSF was integrated into the skin dose tracking system (DTS), a graphical utility for displaying the color-coded skin dose distribution on a 3D graphic of the patient during interventional fluoroscopic procedures. The backscatter convolution method was validated for the non-uniform beam resulting from the use of an ROI attenuator. The ROI attenuator is a copper sheet with about 20% primary transmission (0.7 mm thick) containing a circular aperture; this attenuator is placed in the beam to reduce dose in the periphery while maintaining full dose in the region of interest. The DTS calculated primary plus backscatter distribution is compared to that measured with GafChromic film and that calculated using EGSnrc Monte-Carlo software.

Results:The PSF convolution method used in the DTS software was able to account for the spread of backscatter from the ROI region to the region under the attenuator. The skin dose distribution determined using DTS with the ROI attenuator was in good agreement with the distributions measured with Gafchromic film and determined by Monte Carlo simulation

Conclusion:The PSF convolution technique provides an accurate alternative for entrance skin dose determination with non-uniform primary x-ray beams.