Accuracy of the Small Field Dosimetry Using Acuros XB and AAA Dose Calculation Algorithms of Eclipse Treatment Planning System Within and Beyond Heterogeneous Media for Trubeam 2.0 Unit
G Codel*, E Serin , P Pacaci , E Sanli , M Cebe , H Mabhouti , M Doyuran , E Kucukmorkoc , N Kucuk , A Altinok , D Canoglu , H Acar , H Caglar Ozkok , Medipol University, Istanbul, Istanbul
Presentations
SU-F-SPS-1 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose:: In this study, the comparison of dosimetric accuracy of Acuros XB and AAA algorithms were investigated for small radiation fields incident on homogeneous and heterogeneous geometries
Methods:: Small open fields of Truebeam 2.0 unit (1x1, 2x2, 3x3, 4x4 fields) were used for this study. The fields were incident on homogeneous phantom and in house phantom containing lung, air, and bone inhomogeneities. Using the same film batch, the net OD to dose calibration curve was obtaine dusing Trubeam 2.0 for 6 MV, 6 FFF, 10 MV, 10 FFF, 15 MV energies by delivering 0- 800 cGy. Films were scanned 48 hours after irradiation using an Epson 1000XL flatbed scanner. The dosimetric accuracy of Acuros XB and AAA algorithms in the presence of the inhomogeneities was compared against EBT3 film dosimetry
Results:Open field tests in a homogeneous phantom showed good agreement betweent wo algorithms and measurement. For Acuros XB, minimum gamma analysis passin grates between measured and calculated dose distributions were 99.3% and 98.1% for homogeneousand inhomogeneous fields in thecase of lung and bone respectively. For AAA, minimum gamma analysis passingrates were 99.1% and 96.5% for homogeneous and inhomogeneous fields respectively for all used energies and field sizes.In the case of the air heterogeneity, the differences were larger for both calculations algorithms. Over all, when compared to measurement, theAcuros XB had beter agreement than AAA.
Conclusion:The Acuros XB calculation algorithm in the TPS is an improvemen tover theexisting AAA algorithm. Dose discrepancies were observed for in the presence of air inhomogeneities.