Feasibility Study of Real-Time Ultrasound Monitoring for Abdominal Stereotactic Body Radiation Therapy

no image available
L Su

Lin Su1*, Sook Kien Ng1 , Ying Zhang1 , Tianlong Ji2 , Iulian Iordachita3 , Joseph Herman1 , John Wong1 , H. Tutkun Sen4 , Peter Kazanzides4 , Muyinatu A. Lediju Bell4 , Kai Ding1 , (1) Department of Radiation Oncology, John Hopkins University, Baltimore, MD, (2) Department of Radiation Oncology, The First Hospital of China Medical University, Shenyang, Liaoning, China (3) Department of Mechanical Engineering, Johns Hopkins University, Baltimore, MD, (4) Department of Computer Science, Johns Hopkins University, Baltimore, MD

Presentations

MO-FG-CAMPUS-JeP3-4 (Monday, August 1, 2016) 5:30 PM - 6:00 PM Room: ePoster Theater


Purpose:Ultrasound is ideal for real-time monitoring in radiotherapy with high soft tissue contrast, non-ionization, portability, and cost effectiveness. Few studies investigated clinical application of real-time ultrasound monitoring for abdominal stereotactic body radiation therapy (SBRT). This study aims to demonstrate the feasibility of real-time monitoring of 3D target motion using 4D ultrasound.

Methods:An ultrasound probe holding system was designed to allow clinician to freely move and lock ultrasound probe. For phantom study, an abdominal ultrasound phantom was secured on a 2D programmable respiratory motion stage. One side of the stage was elevated than another side to generate 3D motion. The motion stage made periodic breath-hold movement. Phantom movement tracked by infrared camera was considered as ground truth. For volunteer study, three healthy subjects underwent the same setup for abdominal SBRT with active breath control (ABC). 4D ultrasound B-mode images were acquired for both phantom and volunteers for real-time monitoring. 10 breath-hold cycles were monitored for each experiment. For phantom, the target motion tracked by ultrasound was compared with motion tracked by infrared camera. For healthy volunteers, the reproducibility of ABC breath-hold was evaluated.

Results:Volunteer study showed the ultrasound system fitted well to the clinical SBRT setup. The reproducibility for 10 breath-holds is less than 2 mm in three directions for all three volunteers. For phantom study, the motion between inspiration and expiration captured by camera (ground truth) is 2.35±0.02 mm, 1.28±0.04 mm, 8.85±0.03 mm in LR, AP, SI directly, respectively. The motion monitored by ultrasound is 2.21±0.07 mm, 1.32±0.12mm, 9.10±0.08mm, respectively. The motion monitoring error in any direction is less than 0.5 mm.

Conclusion:The volunteer study proved the clinical feasibility of real-time ultrasound monitoring for abdominal SBRT. The phantom and volunteer ABC studies demonstrated sub-millimeter accuracy of 3D motion movement monitoring.