A Novel Image Analysis Technique (center Pixel Method) to Quantify End-To-End Tests
N Wen1*, K Snyder2 , S Scheib3 , Y Qin4 , H Li5 , I Chetty6 , (1) Henry Ford Health System, Detroit, MI, (2) Henry Ford Hospital System, Detroit, MI, (3) Varian Medical System, Barton, Switzerland, (4) Henry Ford Health System, Detroit, Michigan, (5) Henry Ford Health System, Detroit, Michigan, (6) Henry Ford Health System, Detroit, MI
Presentations
SU-F-J-177 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose:To implement a novel image analysis technique, “center pixel method”, to quantify end-to-end tests accuracy of a frameless, image guided stereotactic radiosurgery system.
Methods:The localization accuracy was determined by delivering radiation to an end-to-end prototype phantom. The phantom was scanned with 0.8 mm slice thickness. The treatment isocenter was placed at the center of the phantom. In the treatment room, CBCT images of the phantom (kVp=77, mAs=1022, slice thickness 1 mm) were acquired to register to the reference CT images. 6D couch correction were applied based on the registration results. Electronic Portal Imaging Device (EPID)-based Winston Lutz (WL) tests were performed to quantify the errors of the targeting accuracy of the system at 15 combinations of gantry, collimator and couch positions. The images were analyzed using two different methods. a) The classic method. The deviation was calculated by measuring the radial distance between the center of the central BB and the full width at half maximum of the radiation field. b) The center pixel method. Since the imager projection offset from the treatment isocenter was known from the IsoCal calibration, the deviation was determined between the center of the BB and the central pixel of the imager panel.
Results:Using the automatic registration method to localize the phantom and the classic method of measuring the deviation of the BB center, the mean and standard deviation of the radial distance was 0.44 ± 0.25, 0.47 ± 0.26, and 0.43 ± 0.13 mm for the jaw, MLC and cone defined field sizes respectively. When the center pixel method was used, the mean and standard deviation was 0.32 ± 0.18, 0.32 ± 0.17, and 0.32 ± 0.19 mm respectively.
Conclusion:Our results demonstrated that the center pixel method accurately analyzes the WL images to evaluate the targeting accuracy of the radiosurgery system.
Funding Support, Disclosures, and Conflict of Interest: The work was supported by a Research Scholar Grant, RSG-15-137-01-CCE from the American Cancer Society.