Quantification of Tracking Performance of a Multi-Layer Electronic Portal Imaging Device

no image available
Y Hu

Y Hu*, J Rottmann , M Myronakis , R Berbeco , Department of Radiation Oncology, Brigham and Women's Hospital, Dana-Farber Cancer Institute and Harvard Medical School, Boston, MA

Presentations

SU-G-BRA-6 (Sunday, July 31, 2016) 4:00 PM - 6:00 PM Room: Ballroom A


Purpose: The purpose of this study was to quantify the improvement in tumor tracking, with and without fiducial markers, afforded by employing a multi-layer (MLI) electronic portal imaging device (EPID) over the current state-of-the-art, single-layer, digital megavolt imager (DMI) architecture.

Methods: An ideal observer signal-to-noise ratio (d’) approach was used to quantify the ability of an MLI EPID and a current, state-of-the-art DMI EPID to track lung tumors from the treatment beam’s-eye-view. Using each detector modulation transfer function (MTF) and noise power spectrum (NPS) as inputs, a detection task was employed with object functions describing simple three-dimensional Cartesian shapes (spheres and cylinders). Marker-less tumor tracking algorithms often use texture discrimination to differentiate benign and malignant tissue. The performance of such algorithms is simulated by employing a discrimination task for the ideal observer, which measures the ability of a system to differentiate two image quantities. These were defined as the measured textures for benign and malignant lung tissue.

Results: The NNPS of the MLI ~25% of that of the DMI at the expense of decreased MTF at intermediate frequencies (0.25≤
Conclusion: MLI performance in tumor tracking is greatly improved by the additional imager layers. This implies that further improvements in tracking may be gained through increasing the thickness of each MLI layer. For tracking, the MLI performance is limited by noise response. Losses in MTF result in negligible differences in d’.

Funding Support, Disclosures, and Conflict of Interest: The project was partially supported by a grant from Varian Medical Systems, Inc. and grant No. R01CA188446-01 from the National Cancer Institute.