Automation of High Dose Rate Brachytherapy Quality Assurance: Development of a Radioluminescent Detection System for Simultaneous Detection of Activity, Timing, and Positioning

no image available
C Jenkins

C Jenkins*, L Xing , B Fahimian , Stanford University, Stanford, CA

Presentations

SU-G-201-3 (Sunday, July 31, 2016) 4:00 PM - 6:00 PM Room: 201


Purpose: Accuracy of positioning, timing and activity is of critical importance for High Dose Rate (HDR) brachytherapy delivery. Respective measurements via film autoradiography, stop-watches and well chambers can be cumbersome, crude or lack dynamic source evaluation capabilities. To address such limitations, a single device radioluminescent detection system enabling automated real-time quantification of activity, position and timing accuracy is presented and experimentally evaluated.

Methods: A radioluminescent sheet was fabricated by mixing Gd₂O₂S:Tb with PDMS and incorporated into a 3D printed device where it was fixated below a CMOS digital camera. An Ir-192 HDR source (VS2000, VariSource iX) with an effective active length of 5 mm was introduced using a 17-gauge stainless steel needle below the sheet. Pixel intensity values for determining activity were taken from an ROI centered on the source location. A calibration curve relating intensity values to activity was generated and used to evaluate automated activity determination with data gathered over 6 weeks. Positioning measurements were performed by integrating images for an entire delivery and fitting peaks to the resulting profile. Timing measurements were performed by evaluating source location and time-stamps from individual images.

Results: Average predicted activity error over 6 weeks was .35 ± .5%. The distance between four dwell positions was determined by the automated system to be 1.99 ± .02 cm. The result from autoradiography was 2.00 ± .03 cm. The system achieved a time resolution of 10 msec and determined the dwell time to be 1.01 sec ± .02 sec.

Conclusion: The system was able to successfully perform automated detection of activity, positioning and timing concurrently under a single setup. Relative to radiochromic and radiographic film-based autoradiography, which can only provide a static evaluation positioning, optical detection of temporary radiation induced luminescence enables dynamic detection of position enabling automated quantification of timing with millisecond accuracy.