no image available
H Kim

H Kim1*, A Chang1 , S Ye2 , (1) Soonchunhyang University Hospital, Seoul, Korea, (2) Seoul National University, Seoul, Korea

Presentations

SU-F-J-176 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:to overcome the several issues of ultrasound image-guided radiation therapy (US IGRT) such as probe pressure and optical tracking disability by using a patient-specific three-dimensional couplant pad (CP) fabricated by a patient’s skin mold using a 3D printing technique.

Methods:A CP was then fabricated by pouring gelatin solution into a fixed-shape container accommodating the patient skin mold fabricated by a 3D printer. A breast phantom was fabricated with the compound of gelatin and agarose and a phantom study was carried out. From four patients who underwent US IGRT, total 486 ultrasound images with and without a CP were acquired before treatment. Effectiveness of the use of the CP was evaluated.

Results:The positioning accuracies in the phantom study were 0.9 ± 0.3 mm and 1.3 ± 0.4 mm with and without the CP in 3D vector amplitude, respectively. In the patient study, the use of CP reduced the mean target shift from 4.7 mm to 3.7 mm in 3D vector amplitude and the one standard deviation from 2.2 mm to 1.7 mm. It also improved the image contrast around the treatment target by 10 %. The centroid offset of the target volume affected from the US scanning coverage and the target deformation due to the excessive probe pressure was decreased from 4.4 mm to 2.9 mm due to the use of CP. Its difference among three different users was statistically significant (p=0.020) without the use of CP but not significantly different (p=0.133) with the use of CP.

Conclusion:Our patient-specific 3D CP using a mold by 3D printing technique is a promising strategy for improving tracking accuracy, image quality, and inter-observer variation for ultrasound-based image guided radiotherapy. In addition to its conventional advantage of non-invasiveness, US can be more facilitated in radiotherapy by the developed CP.