Tumor Tracking in An Iterative Volume of Interest Based 4D CBCT Reconstruction


R Martin

R Martin1*, M Ahmad2 , T Pan1 , (1) UT MD Anderson Cancer Center, Houston, TX, (2) Stanford University, Palo Alto, CA

Presentations

SU-G-BRA-11 (Sunday, July 31, 2016) 4:00 PM - 6:00 PM Room: Ballroom A


Purpose: 4D CBCT can allow evaluation of tumor motion immediately prior to radiation therapy, but suffers from heavy artifacts that limit its ability to track tumors. Various iterative and compressed sensing reconstructions have been proposed to reduce these artifacts, but are costly time-wise and can degrade the image quality of bony anatomy for alignment with regularization. We have previously proposed an iterative volume of interest (I4D VOI) method which minimizes reconstruction time and maintains image quality of bony anatomy by focusing a 4D reconstruction within a VOI. The purpose of this study is to test the tumor tracking accuracy of this method compared to existing methods.

Methods: Long scan (8-10 mins) CBCT data with corresponding RPM data was collected for 12 lung cancer patients. The full data set was sorted into 8 phases and reconstructed using FDK cone beam reconstruction to serve as a gold standard. The data was reduced in way that maintains a normal breathing pattern and used to reconstruct 4D images using FDK, low and high regularization TV minimization (λ=2,10), and the proposed I4D VOI method with PTVs used for the VOI. Tumor trajectories were found using rigid registration within the VOI for each reconstruction and compared to the gold standard.

Results: The root mean square error (RMSE) values were 2.70mm for FDK, 2.50mm for low regularization TV, 1.48mm for high regularization TV, and 2.34mm for I4D VOI. Streak artifacts in I4D VOI were reduced compared to FDK and images were less blurred than TV reconstructed images.

Conclusion: I4D VOI performed at least as well as existing methods in tumor tracking, with the exception of high regularization TV minimization. These results along with the reconstruction time and outside VOI image quality advantages suggest I4D VOI to be an improvement over existing methods.