A Statistical Model for Analyzing the Rotational Error of Single Iso-Center Technique


J Chang

J Chang1,2, (1) Dept of Radiation Medicine, Northwell Health, Lake Success, NY, (2) Dept of Radiation Oncology, New York Weill Cornell Medical Ctr, New York, NY

Presentations

MO-FG-CAMPUS-JeP3-1 (Monday, August 1, 2016) 5:30 PM - 6:00 PM Room: ePoster Theater


Purpose: To develop a generalized statistical model that incorporates the treatment uncertainty from the rotational error of single iso-center technique, and calculate the additional PTV (planning target volume) margin required to compensate for this error.

Methods: The random vectors for setup and additional rotation errors in the three-dimensional (3D) patient coordinate system were assumed to follow the 3D independent normal distribution with zero mean, and standard deviations σx , σy, σz for setup error and a uniform σR for rotational error. Both random vectors were summed, normalized and transformed to the spherical coordinates to derive the chi distribution with 3 degrees of freedom for the radical distance ρ. PTV margin was determined using the critical value of this distribution for 0.05 significant level so that 95% of the time the treatment target would be covered by ρ. The additional PTV margin required to compensate for the rotational error was calculated as a function of σx , σy, σz and σR.

Results: The effect of the rotational error is more pronounced for treatments that requires high accuracy/precision like stereotactic radiosurgery (SRS) or stereotactic body radiotherapy (SBRT). With a uniform 2mm PTV margin (or σx =σy=σz=0.7mm), a σR=0.32mm will decrease the PTV coverage from 95% to 90% of the time, or an additional 0.2mm PTV margin is needed to prevent this loss of coverage. If we choose 0.2 mm as the threshold, any σR>0.3mm will lead to an additional PTV margin that cannot be ignored, and the maximal σR that can be ignored is 0.0064 rad (or 0.37°) for iso-to-target distance=5cm, or 0.0032 rad (or 0.18°) for iso-to-target distance=10cm.

Conclusions: The rotational error cannot be ignored for high-accuracy/-precision treatments like SRS/SBRT, particularly when the distance between the iso-center and target is large.
.