Implementation of Live EPID-Based Inspiration Level Assessment (LEILA) for Deepinspiration Breath-Hold (DIBH) Monitoring Using MV Fluoroscopy


J Lehmann

J Lehmann1,2,3*, J Sun1 , M Doebrich1,3 , B Zwan3,4 , T Fuangrod1 , S Bhatia1 , P Greer1,3 , (1) Calvary Mater Newcastle, Newcastle, Australia (2) The University of Sydney, Sydney, Australia (3) The University of Newcastle, Newcastle, Australia (4) Central Coast Cancer Centre, Gosford, Australia

Presentations

TH-CD-207A-2 (Thursday, August 4, 2016) 10:00 AM - 12:00 PM Room: 207A


Purpose:
As prior work has shown that current DIBH monitoring approaches using surrogate measures (marker block on chest) do not always correspond with the clinical quantity of interest (lung depth, LD), a software tool and workflow are introduced to use MV fluoroscopy during treatment for real-time / Live EPID-based Inspiration Level Assessment (LEILA).

Methods:
A prototype software tool calculates and displays the LD during the treatment of left sided breast cancer. Calculations are based on MV cine images which are acquired with the treatment beam thereby not incurring any additional imaging dose. Image capture and processing are implemented using a dedicated frame grabber computer. The calculation engine automatically detects image orientation and includes provisions for large treatment fields that exceed the size of the EPID panel. LD is measured along a line profile in the middle of the field. LEILA’s interface displays the current MV image, a reference image (DRR), the current LD, as well as a trace of LD over treatment time. The display includes patient specific LD tolerances. Tolerances are specified for each field and loaded before the treatment. A visual warning is generated when the tolerance is exceeded. LEILA is initially run in parallel with current DIBH techniques. When later run by itself DIBH setup will be done using skin marks and room laser.

Results:
Offline tests of LEILA confirmed accurate automatic LD measurement for a variety of patient geometries. Deployment of the EPID during all left sided breast treatments was well tolerated by patients and staff during a multi-month pilot. The frame grabber provides 11 frames-per-second; the MATLAB based LEILA prototype software can analyze five frames-per-second standalone on standard desktop hardware.

Conclusion:
LEILA provides an automated approach to quantitatively monitor LD on MV images during DIBH treatment. Future improvements include a database and further speed optimization.