Prostate Brachytherapy Activity Measurement and Source Localization by Using a Dual Photon Emission Computed Tomography System: A Feasibility Study

no image available
C Lin

C Lin1*, Y Tsai2 , H Lin1, K Chuang1 , C Chiang1 , (1) National Tsing Hua University, Hsinchu, Taiwan, (2) National Taiwan University Hospital, Taipei City, Taiwan

Presentations

SU-G-IeP4-14 (Sunday, July 31, 2016) 5:30 PM - 6:00 PM Room: ePoster Theater


Purpose:To monitor the activity distribution and needle position during and after implantation in operating rooms.

Methods: Simulation studies were conducted to assess the feasibility of measurement activity distribution and seed localization using the DuPECT system. The system consists of a LaBr3-based probe and planar detection heads, a collimation system, and a coincidence circuit. The two heads can be manipulated independently. Simplified Yb-169 brachytherapy seeds were used. A water-filled cylindrical phantom with a 40-mm diameter and 40-mm length was used to model a simplified prostate of the Asian man. Two simplified seeds were placed at a radial distance of 10 mm and tangential distance of 10 mm from the center of the phantom. The probe head was arranged perpendicular to the planar head. Results of various imaging durations were analyzed and the accuracy of the seed localization was assessed by calculating the centroid of the seed.

Results: The reconstructed images indicate that the DuPECT can measure the activity distribution and locate the seeds dwelt in different positions intraoperatively. The calculated centroid on average turned out to be accurate within the pixel size of 0.5 mm. The two sources were identified when the duration is longer than 15 s. The sensitivity measured in water was merely 0.07 cps/MBq.

Conclusion: Preliminary results show that the measurement of the activity distribution and seed localization are feasible using the DuPECT system intraoperatively. It indicates the DuPECT system has potential to be an approach for dose-distribution-validation. The efficacy of acvtivity distribution measurement and source localization using the DuPECT system will evaluated in more realistic phantom studies (e.g., various attenuation materials and greater number of seeds) in the future investigation.