Fluoroscopic X-Ray Beam Profiles for Spectra Incorporating Copper Filtration
K Wunderle1,2*, A Godley1 , Z Shen1 , J Rakowski2 , F Dong1 , (1) Cleveland Clinic Foundation, Cleveland, OH, (2) Wayne State University School of Medicine, Detroit, MI
Presentations
SU-F-I-76 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose:The purpose of this investigation is to characterize and quantify X-ray beam profiles for fluoroscopic x-ray beam spectra incorporating spectral (copper) filtration.
Methods:A PTW (Freiburg, Germany) type 60016 silicon diode detector and PTW MP3 water tank were used to measure X-ray beam profiles for 60, 80, 100 and 120 kVp x-ray beams at five different copper filtration thicknesses ranging from 0-0.9 mm at 22 and 42 cm fields of view and depths of 1, 5, and 10 cm in both the anode-cathode axis (inplane) and cross-plane directions. All measurements were acquired on a Siemens (Erlangen, Germany) Artis ZeeGo fluoroscope inverted from the typical orientation providing an x-ray beam originating from above the water surface with the water level set at 60 cm from the focal spot.
Results:X-ray beam profiles for beam spectra without copper filtration compared well to previously published data by Fetterly et al. [Med Phys, 28, 205 (2001)]. Our data collection benefited from the geometric orientation of the fluoroscope, providing a beam perpendicular to the tank water surface, rather than through a thin side wall as did the previously mentioned study. Profiles for beams with copper filtration were obtained which have not been previously investigated and published. Beam profiles in the anode-cathode axis near the surface and at lower x-ray energy exhibited substantial heel effect, which became less pronounced at greater depth. At higher energy with copper filtration in the beam, the dose falloff out-of-field became less pronounced, as would be anticipated given higher scatter photon energy.
Conclusion:The x-ray beam profile data for the fluoroscopic x-ray beams incorporating copper filtration are intended for use as reference data for estimating doses to organs or soft tissue, including fetal dose, involving similar beam qualities or for comparison with mathematical models.