Ultrasound Imaging of Absorbable Inferior Vena Cava Filters for Proper Placement
T Mitcham1 , R Bouchard1 , A Melancon1 , M Eggers2 , M Melancon1*, (1) University of Texas MD Anderson Cancer Center, Houston, TX, (2) Adient Medical Technologies, Pearland, TX,
Presentations
SU-G-IeP4-15 (Sunday, July 31, 2016) 5:30 PM - 6:00 PM Room: ePoster Theater
Purpose:
Inferior vena cava filters (IVCFs) are used in patients with a high risk of pulmonary embolism in situations when the use of blood thinning drugs would be inappropriate. These filters are implanted under x-ray guidance; however, this provides a dose of ionizing radiation to both patient and physician. B-mode ultrasound (US) imaging allows for localization of certain implanted devices without radiation dose concerns. The goal of this study was to investigate the feasibility of imaging the placement of absorbable IVCFs using US imaging to alleviate the dosage concern inherent to fluoroscopy.
Methods:
A phantom was constructed to mimic a human IVC using tissue-mimicking material with 0.5 dB/cm/MHz acoustic attenuation, while agar inclusions were used to model acoustic mismatch at the venous interface. Absorbable IVCF’s were imaged at 15 cm depth using B-mode US at 2, 3, 5, and 7 MHz transmit frequencies. Then, to determine temporal stability, the IVCF was left in the phantom for 10 weeks; during this time, the IVCF was imaged using the same techniques as above, while the integrity of the filter was analyzed by inspecting for fiber discontinuities.
Results:
Visualization of the inferior vena cava filter was possible at 5, 7.5, and 15 cm depth at US central frequencies of 2, 3, 5, and 7 MHz. Imaging the IVCF at 5 MHz yielded the clearest images while maintaining acceptable spatial resolution for identifying the IVCF’s, while lower frequencies provided noticeably worse image quality. No obvious degradation was observed over the course of the 10 weeks in a static phantom environment.
Conclusion:
Biodegradable IVCF localization was possible up to 15 cm in depth using conventional B-mode US in a tissue-mimicking phantom. This leads to the potential for using B-mode US to guide the placement of the IVCF upon deployment by the interventional radiologist.
Funding Support, Disclosures, and Conflict of Interest: Mitch Eggers is an owner of Adient Medical Technologies. There are no other conflicts of interest to disclose.