Evaluation of the MaxFOV Extended Field of View (EFOV) Reconstruction Algorithm On a GE RT590 CT Scanner


S Grzetic

S Grzetic*, M Weldon , K Noa , N Gupta , Ohio State University Columbus, OH

Presentations

SU-G-TeP2-9 (Sunday, July 31, 2016) 4:30 PM - 5:00 PM Room: ePoster Theater


Purpose: This study compares the newly released MaxFOV Revision 1 EFOV reconstruction algorithm for GE RT590 to the older WideView EFOV algorithm. Two radiotherapy overlays from Q-fix and Diacor, are included in our analysis. Hounsfield Units (HU) generated with the WideView algorithm varied in the extended field (beyond 50cm) and the scanned object’s border varied from slice to slice. A validation of HU consistency between the two reconstruction algorithms is performed.

Methods: A CatPhan 504 and CIRS062 Electron Density Phantom were scanned on a GE RT590 CT-Simulator. The phantoms were positioned in multiple locations within the scan field of view so some of the density plugs were outside the 50cm reconstruction circle. Images were reconstructed using both the WideView and MaxFOV algorithms. The HU for each scan were characterized both in average over a volume and in profile.

Results: HU values are consistent between the two algorithms. Low-density material will have a slight increase in HU value and high-density material will have a slight decrease in HU value as the distance from the sweet spot increases. Border inconsistencies and shading artifacts are still present with the MaxFOV reconstruction on the Q-fix overlay but not the Diacor overlay (It should be noted that the Q-fix overlay is not currently GE-certified). HU values for water outside the 50cm FOV are within 40HU of reconstructions at the sweet spot of the scanner. CatPhan HU profiles show improvement with the MaxFOV algorithm as it approaches the scanner edge.

Conclusion: The new MaxFOV algorithm improves the contour border for objects outside of the standard FOV when using a GE-approved tabletop. Air cavities outside of the standard FOV create inconsistent object borders. HU consistency is within GE specifications and the accuracy of the phantom edge improves. Further adjustments to the algorithm are being investigated by GE.