Clinical Assessment of Auto Beam-Hold Triggered by Fiducial Localization During Prostate RapidArc Delivery


P Atkinson

P Atkinson*, Q Chen , Flower Hospital, Sylvania, OH

Presentations

SU-F-P-30 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To assess the clinical efficacy of auto beam hold during prostate RapidArc delivery, triggered by fiducial localization on kV imaging with a Varian True Beam.

Methods: Prostate patients with four gold fiducials were candidates in this study. Daily setup was accomplished by aligning to fiducials using orthogonal kV imaging. During RapidArc delivery, a kV image was automatically acquired with a momentary beam hold every 60 degrees of gantry rotation. The position of each fiducial was identified by a search algorithm and compared to a predetermined 1.4 cm diameter target area. Treatment continued if all the fiducials were within the target area. If any fiducial was outside the target area the beam hold was not released, and the operators determined if the patient needed re-alignment using the daily setup method.

Results: Four patients were initially selected. For three patients, the auto beam hold performed seamlessly. In one instance, the system correctly identified misaligned fiducials, stopped treatment, and the patient was re-positioned. The fourth patient had a prosthetic hip which sometimes blocked the fiducials and caused the fiducial search algorithm to fail. The auto beam hold was disabled for this patient and the therapists manually monitored the fiducial positions during treatment. Average delivery time for a 2-arc fraction was increased by 59 seconds. Phantom studies indicated the dose discrepancy related to multiple beam holds is <0.1%. For a plan with 43 fractions, the additional imaging increased dose by an estimated 68 cGy.

Conclusions: Automated intrafraction kV imaging can effectively perform auto beam holds due to patient movement, with the exception of prosthetic hip patients. The additional imaging dose and delivery time are clinically acceptable. It may be a cost-effective alternative to Calypso in RapidArc prostate patient delivery. Further study is warranted to explore its feasibility under various clinical conditions.