Spatial and Rotational Quality Assurance of 6DOF Patient Tracking Systems
AH Belcher*, X Liu , Z Grelewicz , R Wiersma , The University of Chicago, Chicago, IL
Presentations
TH-AB-202-11 (Thursday, August 4, 2016) 7:30 AM - 9:30 AM Room: 202
Purpose: External tracking systems used for patient positioning and motion monitoring during radiotherapy are now capable of detecting both translations and rotations (6DOF). In this work, we develop a novel technique to evaluate the 6DOF performance of external motion tracking systems. We apply this methodology to an infrared (IR) marker tracking system and two 3D optical surface mapping systems in a common tumor 6DOF workspace.
Methods: An in-house designed and built 6DOF parallel kinematics robotic motion phantom was used to follow input trajectories with sub-millimeter and sub-degree accuracy. The 6DOF positions of the robotic system were then tracked and recorded independently by three optical camera systems. A calibration methodology which associates the motion phantom and camera coordinate frames was first employed, followed by a comprehensive 6DOF trajectory evaluation, which spanned a full range of positions and orientations in a 20x20x16 mm and 5x5x5 degree workspace. The intended input motions were compared to the calibrated 6DOF measured points.
Results: The technique found the accuracy of the IR marker tracking system to have maximal root mean square error (RMSE) values of 0.25 mm translationally and 0.09 degrees rotationally, in any one axis, comparing intended 6DOF positions to positions measured by the IR camera. The 6DOF RSME discrepancy for the first 3D optical surface tracking unit yielded maximal values of 0.60 mm and 0.11 degrees over the same 6DOF volume. An earlier generation 3D optical surface tracker was observed to have worse tracking capabilities than both the IR camera unit and the newer 3D surface tracking system with maximal RMSE of 0.74 mm and 0.28 degrees within the same 6DOF evaluation space.
Conclusion: The proposed technique was effective at evaluating the performance of 6DOF patient tracking systems. All systems examined exhibited tracking capabilities at the sub-millimeter and sub-degree level within a 6DOF workspace.