Program Information
Position Monitoring for Intracranial SRS Using BrainLAB ExacTrac Snap Verification
S Jang , T McCaw*, M Huq , University of Pittsburgh Cancer Institute, Pittsburgh, PA
Presentations
SU-F-J-25 (Sunday, July 31, 2016) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose: To determine the accuracy of position monitoring with BrainLAB ExacTrac snap verification following couch rotations during intracranial SRS.
Methods: A CT scan of an anthropomorphic head phantom was acquired using 1.25mm slices. The isocenter was positioned near the centroid of the frontal lobe. The head phantom was initially aligned on the treatment couch using cone-beam CT, then repositioned using ExacTrac x-ray verification with residual errors less than 0.2mm and 0.2°. Snap verification was performed over the full range of couch angles in 15° increments with known positioning offsets of 0-3mm applied to the phantom along each axis. At each couch angle, the smallest tolerance was determined for which no positioning deviation was detected.
Results: For couch angles 30°-60° from the center position, where the longitudinal axis of the phantom is approximately aligned with the beam axis of one x-ray tube, snap verification consistently detected positioning errors exceeding the maximum 8mm tolerance. Defining localization error as the difference between the known offset and the minimum tolerance for which no deviation was detected, the RMS error is mostly less than 1mm outside of couch angles 30°-60° from the central couch position. Given separate measurements of patient position from the two imagers, whether to proceed with treatment can be determined by the criterion of a reading within tolerance from just one (OR criterion) or both (AND criterion) imagers. Using a positioning tolerance of 1.5mm, snap verification has sensitivity and specificity of 94% and 75%, respectively, with the AND criterion, and 67% and 93%, respectively, with the OR criterion. If readings exceeding maximum tolerance are excluded, the sensitivity and specificity are 88% and 86%, respectively, with the AND criterion.
Conclusion: With a positioning tolerance of 1.5mm, ExacTrac snap verification can be used during intracranial SRS with sensitivity and specificity between 85% and 90%.
Contact Email: