Dosimetric Validation of AcurosXB Algorithm in Comparison with AAA & CCC Algorithms for VMAT Technique
M Kathirvel*, V Sai subramanian, G Arun, S Thirumalaiswamy, K Ramalingam, S Ashok kumar, K Jagadeesh, Yashoda Hospitals, Hyderabad, Andhra Pradesh, India
SU-E-T-516 Sunday 3:00:00 PM - 6:00:00 PM Room: Exhibit HallPurpose: To dosimetrically validate AcurosXB algorithm for Volumetric Modulated Arc Therapy (VMAT) in comparison with standard clinical Anisotropic Analytic Algorithm(AAA) and Collapsed Cone Convolution(CCC) dose calculation algorithms.
Methods: AcurosXB dose calculation algorithm is available with Varian Eclipse treatment planning system (V10). It uses grid-based Boltzmann equation solver to predict dose precisely in lesser time. This study was made to realize algorithms ability to predict dose accurately as its delivery for which five clinical cases each of Brain, Head&Neck, Thoracic, Pelvic and SBRT were taken. Verification plans were created on multicube phantom with iMatrixx-2D detector array and then dose prediction was done with AcurosXB, AAA & CCC (COMPASS System) algorithm and the same were delivered onto CLINAC-iX treatment machine. Delivered dose was captured in iMatrixx plane for all 25 plans. Measured dose was taken as reference to quantify the agreement between AcurosXB calculation algorithm against previously validated AAA and CCC algorithm. Gamma evaluation was performed with clinical criteria distance-to-agreement 3&2mm and dose difference 3&2% in omnipro-I'MRT software. Plans were evaluated in terms of correlation coefficient, quantitative area gamma and average gamma.
Results: Study shows good agreement between mean correlation 0.9979±0.0012, 0.9984±0.0009 & 0.9979±0.0011 for AAA, CCC & Acuros respectively. Mean area gamma for criteria 3mm/3% was found to be 98.80±1.04, 98.14±2.31, 98.08±2.01 and 2mm/2% was found to be 93.94±3.83, 87.17±10.54 & 92.36±5.46 for AAA, CCC & Acuros respectively. Mean average gamma for 3mm/3% was 0.26±0.07, 0.42±0.08, 0.28±0.09 and 2mm/2% was found to be 0.39±0.10, 0.64±0.11, 0.42±0.13 for AAA, CCC & Acuros respectively.
Conclusions: This study demonstrated that the AcurosXB algorithm had a good agreement with the AAA & CCC in terms of dose prediction. In conclusion AcurosXB algorithm provides a valid, accurate and speedy alternative to AAA and CCC algorithms in a busy clinical environment.