Motion Adaptive Image Filter for Low Dose X-Ray Fluoroscopy in the Real-Time Tumor-Tracking Radiotherapy System

no image available
N Miyamoto

N Miyamoto*, M Ishikawa, K Sutherland, R Suzuki, T Matsuura, S Takao, C Toramatsu, H Nihongi, S Shimizu, R Onimaru, K Umegaki, H Shirato, Graduate School of Medicine, Hokkaido University,Sapporo, Japan

SU-E-J-42 Sunday 3:00:00 PM - 6:00:00 PM Room: Exhibit Hall

Purpose:
In the real-time tumor-tracking radiotherapy system, fiducial markers are detected by X-ray fluoroscopy. The fluoroscopic parameters should be optimized as low as possible in order to reduce unnecessary imaging dose. However, the fiducial markers could not be recognized due to effect of statistical noise in low dose imaging. Image processing is envisioned to be a solution to improve image quality and to maintain tracking accuracy. In this study, a recursive image filter adapted to target motion is proposed.

Methods:
A fluoroscopy system was used for the experiment. A spherical gold marker was used as a fiducial marker. About 450 fluoroscopic images of the marker were recorded. In order to mimic respiratory motion of the marker, the images were shifted sequentially. The tube voltage, current and exposure duration were fixed at 65 kV, 50 mA and 2.5 msec as low dose imaging condition, respectively. The tube current was 100 mA as high dose imaging. A pattern recognition score (PRS) ranging from 0 to 100 and image registration error were investigated by performing template pattern matching to each sequential image. The results with and without image processing were compared.

Results:
In low dose imaging, the image registration error and the PRS without the image processing were 2.15±1.21 pixel and 46.67±6.40, respectively. Those with the image processing were 1.48±0.82 pixel and 67.80±4.51, respectively. There was no significant difference in the image registration error and the PRS between the results of low dose imaging with the image processing and that of high dose imaging without the image processing.

Conclusions:
The results showed that the recursive filter was effective in order to maintain marker tracking stability and accuracy in low dose fluoroscopy.