Encrypted login | home

Program Information

Radiation Dosimetry of 99mTc-IDA-D-[c(RGDfK)]2, a SPECT Agent for Angiogenesis Imaging

no image available
J Kim

J Kim1*, (1) Korea Research Institute of Standards and Science, Daejeon

Presentations

MO-FG-CAMPUS-I-15 (Monday, July 13, 2015) 5:30 PM - 6:00 PM Room: Exhibit Hall


Purpose: Tc-99m labeled IDA-D-[c(RGDfK)]2 (99mTc-RGD) is a recently developed radiotracer for gamma camera or single photon emission computed tomography (SPECT) imaging and promising agent for the visualization of angiogenesis. In this study, we investigated the internal radiation dosimetry of 99mTc-RGD in humans.

Methods: Six normal controls (F:M=4:2; 68.3±3.2 years; 56.5±10.7 kg) were participated in this study. Simultaneous anterior and posterior scans of whole-body were performed using dual head gamma camera system. Before the emission scan, transmission scan was performed just before injection of 99mTc-RGD using Co-57 flood source. After an intravenous injection of 388.7±29.3 MBq of 99mTc-RGD, six serial emission scans were performed at 0, 1, 2, 4, 8 and 24 hours post-injection. The anterior and posterior images were geometrically averaged and attenuation correction was applied using transmission scan image. Regions of interest (ROIs) were drawn on liver, gallbladder, kidneys, urinary bladder, spleen, brain, and large intestine. Time activity curves were obtained from serial emission scan and ROIs. The number of disintegrations per unit activity administered (residence time) were calculated from the area under the curve of time activity curves and injected dose of each patient. Finally, the radiation dose for each organ and effective doses were obtained using OLINDA/EXM 1.1 software and residence time.

Results: High radiation doses were reported on renal and biliary excretion tracks such as urinary bladder wall, upper large intestine, kidneys, liver and gallbladder wall and their doses were 19.15±6.84, 19.28±4.78, 15.67±0.90, 9.13±1.71 and 9.09±2.03 μGy/MBq, respectively. The effective dose and effective dose equivalent were 5.08±0.53 and 7.11±0.58 μSv/MBq, respectively.

Conclusion: We evaluated the radiation dose of 99mTc-RGD, which has an acceptable effective radiation dose compare to the other Tc-99m labeled radio-tracers.


Contact Email: