Inference of Prostate PTV Margins in VMAT Delivery From Intra-Fraction Prostate Motion During SBRT Delivery

no image available
K Thind

K Thind1*, R Wong2 , D Wong3 , C Gerdes4 , T Chow5 , (1) Juravinski Cancer Centre, Hamilton, Ontario, (2) Juravinski Cancer Centre, Hamilton, Ontario, (3) McMaster University, Hamilton, Ontario, (4) Juravinski Cancer Centre, Hamilton, Ontario, (5) Juravinski Cancer Centre, Hamilton, Ontario

Presentations

TU-G-CAMPUS-J-1 (Tuesday, July 14, 2015) 4:30 PM - 5:00 PM Room: Exhibit Hall


Purpose: To retrospectively quantify the intra-fraction prostate motion during stereotactic body radiation therapy (SBRT) treatment using CyberKnife’s target tracking system, which may provide insight into expansion margins from GTV to PTV used in gantry-based treatments. CyberKnife is equipped with an active tracking system (InTempo) that tracks the four fiducials placed in the prostate gland. The system acquires intra-fraction orthogonal kV images at 45° and 315° in a sequential fashion.

Methods: A total of 38 patients treated with SBRT using CyberKnife between 2011 and 2013 were studied. Dose-regime was 36.25 Gy in 5 fractions (7.25 Gy/fraction, twice per week) as per RTOG 0938 guidelines. The CyberKnife image tracking logs for all SBRT treatments using InTempo were examined. A total of 13663 images were examined for the superior/inferior (SI), anterior/posterior (AP) and left/right (LR) translation as well as roll, pitch and yaw rotations for the target position relative to the last known model position.

Results: The mean ± 2 SD of intra-fraction motion was contained within 3 mm for SI and LR and 4.5 mm for AP directions at 5 minutes into the treatment delivery. It was contained within 4 mm for SI and LR and 5 mm for AP at 10 minutes. At 15 minutes into delivery, all translations were contained within 5 mm. The mean ± 2 SD of prostate roll, pitch and yaw increased with time but were contained within 5 degree at 5, 10 and 15 minutes into treatment. Additionally, target translations and rotations were within ± 1 mm and ± 1 degree for 90% and 78% of the time.

Conclusion: The organ motion component of PTV margin for 10 minute VMAT delivery is contained within 4 mm in SI and LR direction and within 5 mm in the AP direction.