Purpose:
To analyze the radiation transmission of the Agility MLC and make recommendation based on the MLC dosimetric characteristics for SRT, SBRT and VMAT planning
Method and Materials:
Agility MLC is the newest model from Elekta and has no back up diaphragm behind leaves for this generation. Leaves are single focused with rounded end; composed of leafs each 0.5cm wide, 9cm thick, constructed from tungsten alloy and provide low transmission <0.5%. Total radiation transmission from leaves and diaphragm is <0.13%. A 360degree arc was generated using iCom tools; leaves were programmed closed while keeping the diaphragm fully open to maximize the MLC transmission effect. Gafchromic EBT films were sandwiched between 4cm of solid water and situated at midplane to take dose measurement. 5000MU was delivered using 6MV VersaHD, various collimator angles, and a 5cm central axis offset was tested also. Films were scanned with Epson 10000XL scanner and analyzed using DoseLab Pro.
Results:
Due to the rounded leaf end and nature of rotation therapy, dose accumulation through the leaf gap is significant. By offsetting the leaf gap from central axis, this accumulation can be greatly reduced. There are dark bands showing accumulation of interleaf transmission which is improved by increasing collimator angle from 0 to 45dgree. However for 45 degree, in most cases, there are larger volumes sweeping under MLC alone, which needs considered planning.
Conclusions:
While inter-leaf leakage is minimized by using collimator angles greater than 0 degrees, the location of the leaf gap must also be managed. The leaf gap position becomes critically important when the treatment area is off axis such is the case when more than one PTV is being treated. With VMAT for SRT, SBRT becoming a more popular planning technique, special attention needs to be paid when initially setting up the field geometry.