Program Information
Examining Acquisition Rate for Using MatriXX Ion Chamber Array to Measure HDR Brachytherapy Treatments
M Wagar*, M Bhagwat , D O'Farrell , S Friesen , I Buzurovic , A Damato , P Devlin , R Cormack , Dana-Farber/Brigham and Women's Cancer Center, Harvard Medical School, Boston, MA
Presentations
SU-E-T-413 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose:There are unique obstacles to implementing the MatriXX ionchamber array as a QA tool in Brachytherapy given that the device is designed for use in the MV energy range. One of the challenges we investigate is the affect of acquisition rates on dose measurement accuracy for HDR treatment plans.
Methods:A treatment plan was optimized in Oncentra Brachy TPS to deliver a planar dose to a 5x5cm region at 10mm depth. The applicator was affixed to the surface of the MatriXX array. The plan was delivered multiple times using a Nucleatron HDR afterloader with a 2.9Ci Ir192 source. For each measurement the sampling rate of the MatriXX movie mode was varied (30ms and 500ms). This experiment was repeated with identical parameters, following a source exchange, with an 11.2Ci Ir192 source. Finally, a single snap measurement was acquired. Analysis was preformed to evaluate the fidelity of the dose delivery for each iteration of the experiment. Evaluation was based on the comparison between the measured and TPS predicted dose.
Results:Higher sample rates induce a greater discrepancy between the predicted and measured dose. Delivering the plan using a lower activity source also produced greater discrepancy in the measurement due to the increased delivery time. Analyzing the single snap measurement showed little difference from the 500ms integral dose measurement.
Conclusion:The advantage of using movie mode for HDR treatment delivery QA is the ability for real time source tracking in addition to dose measurement. Our analysis indicates that 500ms is an optimal frame rate.
Contact Email: