Purpose: Absorbable IVC filters are shown to be safe and efficacious in preventing pulmonary embolism. These absorbable filters disappear from the body after their required duration, alleviating costly removal procedures and downstream complications. Monitoring the positioning and integrity of absorbable devices using dual-energy computed tomography (DECT) would improve treatment efficacy. The purpose of this study is to determine the limit of detection and the energy dependence of DECT for various contrast agents that may be infused within the IVC filters including gold nanoparticles (AuNP) having diameters of 2 and 4 nm.

Methods: All imaging studies were performed on a GE Discovery CT750 system in Gemstone Spectral Imaging (GSI) mode. Plastic vials containing the contrast agent solutions of water and blood were placed in a water bath, and images were acquired with the GSI-5 preset. The images were reformatted into the coronal plane and 5mm diameter ROIs were placed within each solution on a GE Advantage Workstation. Monoenergetic reconstructions were generated from 40 – 140 keV.

Results: Mass attenuation (contrast per unit density) for AuNPs was greater than iron, but less than barium and iodine. Contrast was 10.2 (± 3.6) HU for 4 nm AuNP at 0.72 mg/ml and 12.1 (± 4.2) for 2 nm AuNP at 0.31 mg/ml at 70 keV suggesting reasonable chance of visualization at these concentrations for 70 keV reconstruction. The contrast as a function of CT energy is similar in both water and blood. Iodine is most dependent, followed closely by barium and iron, and trailed by a large margin by the AuNP. This was unexpected given Au’s large atomic number and the predominance of photoelectric effect at low energy.

Conclusion: Infusion of IVC filters with AuNP is feasible. Discrimination of AuNP-infused IVC filters from surrounding anatomy warrants further investigation.