Portal Dosimetry Commissioning of Multiple (6) Varian TrueBeam Linacs Equipped with PortalVision DMI MV Imager
M Weldon*, D DiCostanzo , S Grzetic , J Hessler , Ohio State University, Columbus, OH
Presentations
SU-E-T-624 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose: To show that a single model for Portal Domisetry (PD) can be established for beam-matched TrueBeam™ linacs that are equipped with the DMI imager (43x43cm effective area).
Methods: Our department acquired 6 new TrueBeam™s, 4 “Slim” and 2 “Edge” models. The Slims were equipped with 6 and 10MV photons, and the Edges with 6MV. MLCs differed between the Slims and Edges (Millennium 120 vs HD-MLC respectively). PD model was created from data acquired using a single linac (Slim). This includes maximum field size profile, as well as output factors and acquired measured fluence using the DMI imager. All identical linacs were beam-matched, profiles were within 1% at maximum field size at a variety of depths. The profile correction file was generated from 40x40 profile acquired at 5cm depth, 95cm SSD, and was adjusted for deviation at the field edges and corners. The PD model and profile correction was applied to all six TrueBeam™s and imagers. A variety of jaw only and sliding window (SW) MLC test fields, as well as TG-119 and clinical SW and VMAT plans were run on each linac to validate the model.
Results: For 6X and 10X, field by field comparison using 3mm/3% absolute gamma criteria passed 90% or better for all cases. This was also true for composite comparisons of TG-199 and clinical plans, matching our current department criteria.
Conclusion: Using a single model per photon energy for PD for the TrueBeam™ equipped with a DMI imager can produce clinically acceptable results across multiple identical and matched linacs. It is also possible to use the same PD model despite different MLCs. This can save time during commissioning and software updates.