A Comparison of Expected Dwell Times and Dose Variations for N-AMD Patients Treated with An Episcleral Brachytherapy Device

no image available
L Marsteller

R Hamilton1 , P Patel2 , K Balaggan2 , M Restori2 , T Ilginis2 , M Drew3 , M McGovern3 , J Vitali3 , L Marsteller3*, (1) Univ Arizona, Tucson, AZ, (2) Moorfields Eye Hospital, London, United Kingdom, (3) Salutaris Medical Devices, Inc., Tucson, AZ

Presentations

SU-E-T-16 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To evaluate the variations in dwell times and doses expected when using an episcleral brachytherapy device for treatment of neovascular age-related macular degeneration (n-AMD) based on accurate imaging modalities.

Methods: Data from 40 eyes from 40 subjects with known n-AMD acquired through the Distance of Choroid Study (DOCS) conducted at Moorfields Eye Hospital was used to determine the target depth; the distance from the outer scleral surface of the eye, through the choroid, to the apex of the choroidal neovascularization (CNV). Each subject underwent, in triplicate, enhanced-depth Spectral Domain Optical Coherence Tomography (SD-OCT), Swept Source Optical Coherence Tomography, (SS-OCT) and Ocular Ultrasound (O-US). These data are the most comprehensive and accurate measurements of the dimensions of the CNV and adjacent layers of the eye for this cohort of patients. During treatment of n-AMD, patients receive a dose of 24Gy to the apex at the target depth. Using the percentage depth dose for a Sr-90 episcleral brachytherapy device, dwell times and doses to the apex were computed to determine the expected variations.

Results: The mean target depth and the 95% confidence interval (CI) determined by combining O-US with SD-OCT were 1326 (956,1696)μm and with SS-OCT were 1332 (970,1693)μm. The calculated corresponding mean dwell times and 95% (CI) were 334 (223,445)s and 335 (226,445)s for SD-OCT and SS-OCT determined depths, respectively. The corresponding mean apex dose and 95% (CI) were 24 (35.9,18.0)Gy (SD-OCT) and 24 (35.6,18.1)Gy (SS-OCT).

Conclusion: For episcleral brachytherapy treatment of n-AMD, using a patient population average target depth for treatment planning is inadequate, resulting in dose variations of a factor of approximately two over the 95% CI and larger variations for a nontrivial segment of the population. Each patient should have individualized imaging studies to determine the target depth for use in the dwell time calculation.


Funding Support, Disclosures, and Conflict of Interest: Study was sponsored by Salutaris Medical Devices, Ltd., a subsidiary of Salutaris Medical Devices, Inc. Hamilton and Marsteller are founders of Salutaris Medical Devices, Inc. Drew, McGovern and Vitali are minor equity holders in Salutaris Medical Devices, Inc.