Program Information
Designing Low-Z Targets To Enhance Surface Dose: A Monte Carlo Simulation
R Kelly1*, J Robar2 , D Parsons3 , (1) Nova Scotia Cancer Centre, Halifax, NS, (2) Capital District Health Authority, Halifax, NS, (3) Dalhousie University, Halifax, Nova Scotia
Presentations
SU-E-T-244 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall
Purpose:Recent developments in The Varian Truebeam linac platform allows for the introduction of low-Z targets into the beam line for the imaging purposes. We have proposed using a low-Z target for radiation therapy purposes to enhance the surface dose during radiation treatment. The target arm of the Varian Truebeam accelerator consists of multiple targets with are linearly translated into the beam line. We have designed two Low-Z targets made of carbon: 1) a step target consisting of three steps of 15%, 30% and 60% CSDA range for 2.5 MeV electrons Figure 1a; 2) and a ramp target, an incline plane 2cm long with thicknesses ranging from 0% to 60% CSDA range, Figure 1b. The purpose of this work will determine the spectral characteristics of these target designs and determine if they have practical clinical applications for enhancing surface dose.
Methods:To calculate the spectral characteristics of these targets, a standard Monte Carlo model of a Varian Clinac accelerator was used. Simulations were performed with a carbon step target, and a carbon ramp target, located at the same position as the electron foil in the rotating carousel. Simulations were carried out using a 2.5 MeV electron beam.
Results:The step target design produced spectral characteristics which were similar to spectral model using a single disk target of the same thickness. The ramp target provides a means to have positional variation of the spectral components of the beam, however, the electron component as 60% CSDA us much broader than the step target.
Conclusion:The carbon step-target provides a spectral distribution which is similar to a carbon disk of comparable thickness. The spectral distribution from the ramp-target can be modified as a function of position to provide a wide range of low energy electrons for surface dose enhancement.
Contact Email: