Effect of Treatment Table and Immobilization Devices On Surface Dose When Using a GRID Technique

no image available
S Gajdos

S Gajdos*, J Donaghue , Akron General Medical Center, Akron, OH

Presentations

SU-E-T-348 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To determine the increase of surface dose of MLC-designed GRID therapy in the presence of immobilization devices and treatment table.

Methods: To create a GRID field, our facility utilizes an MLC consisting of four millimeter wide leaves. The field is designed to have aperture sizes of 0.8 cm X 0.8 cm with inter-aperture distance of 3.2 cm. Gafchromic EBT3 film was placed between the surface of a solid water phantom and the immobilization device. The treatment table was also present within the beam path. The devices consist of carbon fiber exterior shell. A piece of film was also placed at maximal depth for the photon energy of 10 MV. Image files were converted to dose per a calibration curve based on the selected red channel. The surface dose to maximum dose was established by comparing the ratio of seven centrally located aperture regions-of-interest and four adjacent inter-aperture regions-of-interest were measured with the available software tools.

Results: With no devices present in beam path, the ratio of surface dose to maximum dose was 11.5% ± 0.3% for aperture region and 7.0% ± 0.1% for inter-aperture region. When devices are present, the ratio of surface dose to maximum dose was 45.2% ± 0.5% and 33.8% ± 1.1%, respectively. Due to the presence of devices, the surface dose increases in aperture region by 3.8 times or in the inter-aperture region by 4.7 times.

Conclusion: The purpose of using GRID technique is to deliver a single fractional dose in range of 15-20 Gy to a bulky lesion while also preserving skin tolerance. The increase of surface dose due to devices placed in beam path may increase the chance of skin toxicity in GRID therapy. Care should be used to determine best manageable patient immobilization while considering skin dose especially for posteriorly located lesions.