Program Information
Combining DAO with MV + KV Optimization to Improve Skin Dose Sparing with Real-Time Fluoroscopy
Z Grelewicz*, R Wiersma , The University of Chicago, Chicago, IL
Presentations
WE-AB-303-6 (Wednesday, July 15, 2015) 7:30 AM - 9:30 AM Room: 303
Purpose: Real-time fluoroscopy may allow for improved patient positioning and tumor tracking, particularly in the treatment of lung tumors. In order to mitigate the effects of the imaging dose, previous studies have demonstrated the effect of including both imaging dose and imaging constraints into the inverse treatment planning object function. That method of combined MV+kV optimization may result in plans with treatment beams chosen to allow for more gentle imaging beam-on times. Direct-aperture optimization (DAO) is also known to produce treatment plans with fluence maps more conducive to lower beam-on times. Therefore, in this work we demonstrate the feasibility of a combination of DAO and MV+kV optimization for further optimized real-time kV imaging.
Methods: Therapeutic and imaging beams were modeled in the EGSnrc Monte Carlo environment, and applied to a patient model for a previously treated lung patient to provide dose influence matrices from DOSXYZnrc. An MV + kV IMRT DAO treatment planning system was developed to compare DAO treatment plans with and without MV+kV optimization. The objective function was optimized using simulated annealing. In order to allow for comparisons between different cases of the stochastically optimized plans, the optimization was repeated twenty times.
Results: Across twenty optimizations, combined MV+kV IMRT resulted in an average of 12.8% reduction in peak skin dose. Both non-optimized and MV+kV optimized imaging beams delivered, on average, mean dose of approximately 1 cGy per fraction to the target, with peak doses to target of approximately 6 cGy per fraction.
Conclusion: When using DAO, MV+kV optimization is shown to result in improvements to plan quality in terms of skin dose, when compared to the case of MV optimization with non-optimized kV imaging. The combination of DAO and MV+kV optimization may allow for real-time imaging without excessive imaging dose.
Funding Support, Disclosures, and Conflict of Interest: Financial support for the work has been provided in part by NIH Grant T32 EB002103, ACS RSG-13-313-01-CCE, and NIH S10 RR021039 and P30 CA14599 grants. The contents of this submission do not necessarily represent the official views of any of the supporting organizations.
Contact Email: