no image available
R Mao

R Mao1*, Y Zhang2 , L Ren2 , L Tian1 , R Gao2 , H Ge1 , F Yin2 , (1) The Affiliated Cancer Hospital of Zhengzhou University, Zhengzhou, henan, (2) Duke University Medical Center, Durham, NC,

Presentations

SU-E-T-293 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:To evaluate the dosimetry of microscopic disease (MD) region of lung cancer in stereotactic body radiation therapy (SBRT).

Methods:For simplicity, we assume organ moves along one dimension. The probability distribution function of tumor position was calculated according to the breathing cycle. The dose to the MD region was obtained through accumulating the treatment planning system calculated doses at different positions in a breathing cycle. A phantom experiment was then conducted to validate the calculated results using a motion phantom (The CIRS ‘Dynamic’ Thorax Phantom). The simulated breathing pattern used a cos4(x) curve with an amplitude of 10mm. A 3-D conformal 7-field plan with 6X energy was created and the dose was calculated in the average intensity projection (AIP) simulation CT images. Both films (EBT2) and optically stimulated luminescence (OSL) detectors were inserted in the target of the phantom to measure the dose during radiation delivery (Varian Truebeam) and results were compared to planning dose parameters.

Results: The Gamma analysis (3%/3mm) between measured dose using EBT2 film and calculated dose using AIP was 80.5%, indicating substantial dosimetric differences. While the Gamma analysis (3%/3mm) between measured dose using EBT2 and accumulated dose using 4D-CT was 98.9%, indicating the necessity of dose accumulation using 4D-CT. The measured doses using OSL and theoretically calculated doses using probability distribution function at the corresponding position were comparable.

Conclusion:Use of static dose calculation in the treatment planning system could substantially underestimate the actually delivered dose in the MD region for a moving target.

Funding Support, Disclosures, and Conflict of Interest: Funding Supported by NSFC, No.81372436