LINAC-Based Single Isocenter Frameless SRT for Brain Metastases

no image available
B Liu

B Liu, L Zhang, N Rigor, J Kim. City of Hope Foundation, CA

Presentations

SU-E-T-536 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose:
Single-isocenter Stereotactic Radiotherapy of multiple brain metastases with Varian 21 IX LINAC, using Aktina Pinpoint system for patient setup

Methods:
In 2014, five single-isocenter RapidArc SRT plans were delivered to five patients with 2 to 8 brain metastases using Varian 21 IX. Aktina Pinpoint system was used for setup and 2mm PTV margin were used. CBCT was acquired before and after the beam delivery. The prescription is 2100 cGy in 3 fractions. Eclipse planning system was used for treatment planning. Depending on the number of metastases and their locations, 1 to 5 coplanar or non coplanar arcs were used. Typically, 2 or 3 arcs are used. IMRT QAs were performed by comparing an A1SL ion chamber point dose measurement in solid water phantom to point dose of the plan; also, based on EPID measurement, 3D spatial dose was calculated using DosimetryCheck software package from MathResolutions Inc. The EPID system has an active area of 40cm by 30cm with 1024 by 768 photodiodes, which corresponds to a resolution of 0.4mm by 0.4mm pixel dimension.

Results:
For all the plans, at least 95% PTV coverage was achieved for full prescription dose, with plan normalization > 75%. RTOG conformity indices are less than 1.1 and Paddick gradient indices are less than 4.5. The distance from prescription IDL to 50%IDL increases as the number of metastases increases, and it ranges from 0.6mm to 0.8mm. Treatment time varies from 10mins to 30mins, depending on the number of arcs and if the arcs are coplanar. IMRT QA shows that the ion chamber measurement agree with the eclipse calculation within 3%, and 95% of the points passed Gamma, using 3% dose difference and 3mm DTA

Conclusion:
High quality single isocenter RapidArc SRT plan can be optimized and accurately delivered using Eclipse and Varian 21IX.