Emulation of Yb-169 Gamma-Ray Spectrum Using Metal-Filtered 250 KVp X-Rays for Pre-Clinical Studies of Gold Nanoparticle-Aided Radiation Therapy


F Reynoso

F Reynoso*, S Cho , UT MD Anderson Cancer Center, Houston, TX

Presentations

SU-E-T-359 (Sunday, July 12, 2015) 3:00 PM - 6:00 PM Room: Exhibit Hall


Purpose: To develop an external beam surrogate of the Yb-169 brachytherapy source applying a filter-based spectrum modulation technique to 250 kVp x-rays. In-vitro/vivo studies performed with the modulated 250 kVp beam will help gauge the benefits of implementing gold nanoparticle-aided radiotherapy with the Yb-169 source.

Methods: A previously validated MCNP5 model of the Phillips RT-250 orthovoltage unit was used to obtain the percentage depth dose (PDD) and filtered photon spectra for a variety of filtration and irradiation conditions. Photon spectra were obtained using the average flux F4 tally in air right after all collimation. A 30 x 30 x 30 cm³ water phantom was used to compute the PDD along the central axis (CAX) under the standards conditions of a 10 x 10 cm² field size at 50 cm SSD. Cylindrical cells of 4 cm in diameter and the energy deposition F6 tally were used along the CAX to score the doses down to 20 cm depth. The number of particle history was set to 2 x 10⁸ in order to keep the relative uncertainty within each cell < 0.3%. The secondary electron spectrum within a gold-loaded tissue due to each photon spectrum was also calculated using EGSnrc and compared with that due to Yb-169 gamma rays.

Results: Under the practical constraints for the spectrum modulation task, 250 kVp x-rays filtered by a 0.25 mm Erbium (Er) foil produced the best match with Yb-169 gamma rays, in terms of PDD and, more importantly, secondary electron spectrum.

Conclusion: Modulation of 250kVp x-ray spectrum by an Er-filter was found effective in emulating the gamma ray spectrum of Yb-169. Possible benefits as predicted from the current MC model such as enhanced radiosensitization with the Er-filtered beam (as a surrogate of Yb-169) was confirmed with a separate in-vitro study.

Funding Support, Disclosures, and Conflict of Interest: Supported by DOD/PCRP grant W81XWH-12-1-0198